In order to do internal calibration for TOR-3, do you use Standard Oil? If yes, where can we find this Standard Oil?
- 该话题包含 1个回复,2 人参与,最后由 更新于 1月、 1周前 。
Answers
-
十一月 8, 2025 8:35 上午 by David Allen
Yes, calibration of the TOR-3 device is performed using certified reference (standard) insulating oil according to IEC 60156. Equivalent calibration oil can be purchased from certified laboratory-oil suppliers; specifications are listed in the TOR-3 manual.