The TOR-6 diesel fuel moisture and particles content tester, made by GlobeCore, has been developed for the fast analysis of fuel quality directly at the facility. Diesel fuel has to be really pure for engines and fuel equipment to work properly. It’s been shown that even a tiny bit of moisture or mechanical impurities can lead to performance issues, faster system breakdown, and mechanical failure. TOR-6 helps you figure out two important things: how much water is in the fuel, in percentage or ppm form, and how much solid stuff is in there. The assessment is conducted in accordance with international standards, such as ISO 4406. The rapid analysis facilitates the straightforward determination of the fuel’s suitability for utilisation or the necessity for treatment. The apparatus is characterised by its diminutive size and ease of use, and is equipped with a digital display that facilitates the presentation of the results. It’s really easy to use and you don’t need any special training. The TOR-6 is a real all-rounder, you can use it in laboratories, fleets, at gas stations, and out in the field. The job of the tester is to stop people using fuel that isn’t up to scratch, which makes the engine last longer and cuts maintenance costs. We keep an eye on the diesel fuel to make sure the equipment is running smoothly and safely.
Diesel fuel moisture and particles content tester TOR-6 Specifiacaions
| No. | Parameter | Value |
| Moisture measurement characteristics | ||
| 1 | Operating range, aw (active water) | 0…1 |
| 2 | Precision at 20 °C | ±10 % or ±10 ppm |
| 3 | Temperature measurement | Yes |
| Purity or contamination level characteristics | ||
| 4 | Permissible viscosity range, mm²/sec | 1 to 1,000 |
| 5 | Standards maintained in determining the purity class | ISO 99 (ISO 4406:1999)
SAE (SAE AS 4059) or ISO 87 (ISO 4406:1987) NAS (NAS 1638) |
| 6 | Measuring range | Measuring from ISO 9/8/7 (min) to ISO 25/24/23 (max)
Calibrated within the ISO range from 13/11/10 to 23/21/18 |
| 7 | Precision | ISO class ± 1/2 in calibrated range |
| 8 | Required power, W | 3 |
| 9 | Single phase 50 Hz AC power supply voltage, V | 220 |
| 10 | Dimensions, mm, not more than
– length – width – height |
235 275 370 |
| 11 | Weight, kg, not more than | 10 |