What is the price for TOR-1? What is the precision in PPM? And precision specifically in the range of 10ppm – 50ppm? What is the operation time per each sample? Could you provide data spec for the instrument? Any calibration before samples? What recurring maintenances are required? Thanks

Answers

  • November 5, 2025 at 9:32 am by Brian Allen

    The TOR-1 measures dissolved moisture in transformer oil in ppm. Typical accuracy in the 10-50 ppm range is ±2 ppm or ±5% of reading. Analysis time is about 5-10 minutes per sample. The unit comes with a calibration certificate and technical datasheet. Annual calibration is recommended. Regular maintenance includes cleaning the cell, checking connections, and replacing seals if needed.

Only registered users can answer. No account? Sign up!

Sign up

Sign in

To continue log in with Google.